Avtor/Urednik | Poljak, Mario; Marin, Irena J; Seme, Katja; Brinovec, Vladimir; Matičič, Mojca; Meglič-Volkar, Jelka; Lešničar, Gorazd; Vince, Adriana | |
Naslov | Second-generation hybrid capture test and amplicor monitor test generate highly correlated hepatitis B virus DNA levels | |
Tip | članek | |
Vir | J Virol Methods | |
Vol. in št. | Letnik 97 | |
Leto izdaje | 2001 | |
Obseg | str. 165-9 | |
Jezik | eng | |
Abstrakt | The performance of the Digene Hybrid Capture II HBV DNA Test HC II and the Roche Cobas Amplicor Monitor Test (Cobas-HBV) was evaluated on 252 serum samples. One hundred and seventy-three samples were HBV DNA positive and 75 HBV DNA negative by both assays. Four samples were HBV DNA positive by Cobas-HBV only. Linear regression analysis showed that the HBV DNA concentrations obtained from both assays were significantly related (n=173, r=0.976, p<0.0001). The results of the study show that Hybrid capture II and Cobas-HBV could be used equally in the management for pstients with chronic HBV infection. | |
Deskriptorji | HEPATITIS B VIRUS POLYMERASE CHAIN REACTION DNA, VIRAL HEPATITIS B |