Avtor/Urednik     Kandus, A; Ponikvar, R; Drinovec, J; Kladnik, S; Ivanovich, P
Naslov     Anaphylatoxins C3a and C5a adsorption on acrylonitrile membrane of hollow-fiber and plate dialyzer--an in vivo study
Tip     članek
Vir     Int J Artif Organs
Vol. in št.     Letnik 13, št. 3
Leto izdaje     1990
Obseg     str. 176-80
Jezik     eng
Abstrakt     We studied the adsorption of anaphylatoxins C3a and C5a on acrylonitrile (AN69) hollow-fiber (AN69HF) and plate (AN69P) dialyzers in 8 patients during 4-hour hemodialyses (HD). Blood passed first through a cuprophan dialyzer and then through AN69 dialyzers that were not in contact with dialysis fluid. Plasma C3a and C5a were measured in samples taken from the afferent and efferent blood lines of the acrylonitrile dialyzers at 15, 60 and 240 min. Plasma C3a concentrations decreased significantly in blood that had passed through AN69 dialyzers. This decrease, indicating membrane adsorption, was maximal (by 65 percent in AN69HF and by 59 percent in AN69P) at 15 min and minimal (by 53 percent in AN69HF and by 18 percent in AN69P) at 240 min. The decrease in plasma C5a concentrations was smaller and significant throughout HD only with AN69HF. The amount of C3a adsorbed was at least 45,000 micrograms in AN69HF and 18,000 micrograms in AN69P. These findings demonstrate that acrylonitrile dialyzers adsorb more C3a and C5a than they produce. This membrane adsorption may explain why the increase of plasma C3a and C5a is inhibited during HD.
Deskriptorji     COMPLEMENT 3A
COMPLEMENT 5A
KIDNEY, ARTIFICIAL
MEMBRANES, ARTIFICIAL
ACRYLONITRILE
ADSORPTION
ADULT
AGED
COMPLEMENT ACTIVATION
HEMODIALYSIS
KIDNEY FAILURE, CHRONIC
MIDDLE AGE
TIME FACTORS