Avtor/Urednik     Ho, Anthony K; Sibata, Claudio H; Thomadsen, Bruce R
Naslov     Comparison of TDF and LQ models using the bioeffects algorithm of a treatment planning system
Tip     članek
Vir     Radiol Oncol
Vol. in št.     Letnik 31, št. 3
Leto izdaje     1997
Obseg     str. 315-8
Jezik     eng
Abstrakt     In this study, two examples, first with two-field AP/PA plans and second with four-field plans are evaluated for Co-60 beams using the bioeffects program of the Radiation Oncology Computer System (ROCS) treatment planning system. The bioeffects algorithm enables the summation of two or more treatment plans. Biomodifier tables, which convert the values of dose per fracrtion delivered over a period of time to Time Dose Fractionation (TDF) are included with the software. The biomedifier table is a standard ROCS twodimensional table. By using the linear-quadratic (LQ) model, the biological equivalent dose versus the physical absorbed dose was determined and input as a new biomodifier table. The distribution of TDF values and the biological equivalent dose using the LQ model shows that the LQ model may be a better choice for a bioeffect algorithm. Furthermore, the LQ model may be implemented in the ROCS system.
Deskriptorji     NEOPLASMS
RADIOTHERAPY, COMPUTER-ASSISTED
RADIOTHERAPY PLANNING, COMPUTER-ASSISTED
DOSE-RESPONSE RELATIONSHIP, RADIATION