Avtor/Urednik | Mihelin, Marjan; Liščić, Rajka M | |
Naslov | A custom designed system to measure corticospinal tract jitter: tehnical note | |
Tip | članek | |
Vir | Electroencephalogr Clin Neurophysiol | |
Vol. in št. | Letnik 109 | |
Leto izdaje | 1998 | |
Obseg | str. 194-7 | |
Jezik | eng | |
Abstrakt | Typical latency of an individual limb muscle response to magnetic or electric stimulation of the human cortex is in the range of 10-50 ms. For the latency variability, i.e., jitter studies, a resolution of at least 20 micro s is needed. Commercially available EMG equipment needs cutstom-designed upgrading to allow for such studies. Two solutions were designed: (1) a hardware unit allowing an adjustable delay of data acquisition after the delivered stimuli; and (ii) diverting of the amplified biological signal and the EMG eqipment trigger to an ecternal computer equipped with an analogue-to-diginal conversion (ACD) module. Custom-designed software made fast ADC possible during the whole period of data acquisition. Both concepts were applied to a Vickers medical Mystro electromyograph, and have been successfully used in the Ljubljana (Slovenia) Institute of Clinical Neurophysiology for the last 2 years. | |
Deskriptorji | PYRAMIDAL TRACTS MOTOR CORTEX BRAIN MAPPING ELECTROMYOGRAPHY ELECTRIC STIMULATION ELECTROMAGNETIC FIELDS MICROCOMPUTERS MUSCLE, SKELETAL |