Author/Editor     Trebše, Rihard; Milošev, Ingrid; Kovač, Simon; Mikek, Martin; Pišot, Venčeslav
Title     Poor results from the isoelastic total hip replacement: 14-17-year follow-up of 149 cementless prostheses
Type     članek
Source     Acta Orthop Scand
Vol. and No.     Letnik 76, št. 2
Publication year     2005
Volume     str. 169-76
Language     eng
Abstract     Rackground To overcome the mismatch between a stiff stem and the more elasfic bone, the concept of isoelasticity was introduced in the 1970s. This conecpt was based on the assumption that the implant and the bone should deform as one unit to avoid stress shielding. The Rubert Mathys (RM) cementless total hip replacement (THR) was one of the earliest isoelastic designs. Patients and methods From 1984 to 1987, we performed 149 total hip replacements in 135 patients (92 women, mean age 47 (21-72) years) using third-generation Mathys isoelastic polyacetal stem with stainless-steel heads and polyethylene cementless acetabular cups. 11 patients died before revision or before the examination for this study, and 14 were lost to follow-up. Average follow-up time for the remaining 110 patients was 15 (14-17) years. Results To date, 69 hips (64 patients) have been revised, 5 due to infection. The 10-year survival rate for any reason was 70% (66-74). In the 46 remaining patients (53 hips), the average HHS was 80 (39-100) points. 13 of these were regarded as radiographic failures, with an average HHS of 75 points. lnterpretation The performance of this prosthesis was unacceptably poor. Higher debris pruduction and poor primary fixation are believed to be the main reason for the high failure rate.
Descriptors     HIP PROSTHESIS
STAINLESS STEEL
POLYETHYLENES
PROSTHESIS FAILURE
REOPERATION
HIP JOINT
TREATMENT OUTCOME
SURVIVAL ANALYSIS