Author/Editor     Golež, Ana; Rodi, Zoran
Title     Stimulation of Cauda Equina with a figure-of-eight magnetic coil
Type     članek
Vol. and No.     Letnik 32, št. 6
Publication year     2015
Volume     str. 501-504
ISSN     1537-1603 - Journal of clinical neurophysiology : official publication of the American Electroencephalographic Society
Language     eng
Abstract     Purpose: To record responses in the abductor hallucis muscle after lumbosacral stimulation with a figure-of-eight magnetic coil. To compare magnetic stimulation (MS), electrical stimulation (ES), and the peripheral motor conduction time. Methods: M and F waves and compound muscle action potentials were recorded after L1 MS, S1 MS, and L1 ES. Results: The compound muscle action potential latency on L1 MS was 26.9 +/- 2.6 milliseconds and on L1 ES was 25.6 +/- 2.2 milliseconds, and the peripheral motor conduction time was 27.0 +/- 2.5 milliseconds (analysis of variance P = 0.0008). The compound muscle action potential amplitude was 12 +/- 11% of M-wave on L1 MS, 32 +/- 18% on S1 MS, and 31 +/- 17% on L1 ES. Conclusions: Dorsal MS reliably elicits responses in the abductor hallucis muscle. The latency of compound muscle action potentials on L1 MS is comparable to peripheral motor conduction time. A potential drawback of MS may be too weak stimulation at the proximal end of the cauda equina.
Descriptors     Cauda equina
Transcutaneous electric nerve stimulation
Neural conduction
Magnetics
Kavda ekvina
Transkutana električna stimulacija živcev
Prevajanje po živcu
Magnetizem
Physiology
Patophysiology
Fiziologija
Patofiziologija