Author/Editor     el-Khatib, EE; Podgoršak, EB; Pla, C
Title     Broad beam and narrow beam attenuation in Lipowitz's metal
Type     članek
Source     Med Phys
Vol. and No.     Letnik 14, št. 1
Publication year     1987
Volume     str. 135-9
Language     eng
Abstract     Attenuation properties of Lipowitz's metal have been studied for narrow and broad beams of cobalt-60 gamma rays and 4-10 MV x-rays. The measured transmitted fraction for geometries used in radiotherapy depends on the field size and depth of measurement. Therefore a calculation of dose for partially attenuated beams based on narrow beam attenuation coefficients can cause large errors in dosimetry. Our simple calculation of transmitted fractions based on primary attenuation and scattered radiation agrees quite well with the measured data for therapeutic geometries. Also given is a table for linear, mass attenuation, and mass energy absorption coefficients of Lipowitz's metal in the photon energy range from 10 keV to 10 MeV.
Descriptors     ALLOYS
RADIATION PROTECTION
BIOPHYSICS
MODELS, ANATOMIC
RADIOTHERAPY DOSAGE
RADIOTHERAPY, HIGH-ENERGY