Author/Editor | Podgoršak, EB; Evans, MD | |
Title | An endocavitary rectal irradiation technique | |
Type | članek | |
Source | Int J Radiat Oncol Biol Phys | |
Vol. and No. | Letnik 13, št. 12 | |
Publication year | 1987 | |
Volume | str. 1937-41 | |
Language | eng | |
Abstract | An alternative to standard short target-skin distance (TSD) rectal irradiation techniques is presented. The proctoscopic rectal cone is linked outside of the patient to a superficial X ray tube operated at 50 kVp and 25 mA. The cone is rigidly held in place by a hydraulic clamp and linked to the X ray tube with an electromagnetic lock. The X ray beam is filtered by 0.3 mm of aluminum, giving a half-value-layer of 0.27 mm of aluminum and an exposure rate of 720 R/min. (0.186 C/kg-min.) at the treatment TSD of 22 cm. The physical aspects of the technique are described in detail. | |
Descriptors | BRACHYTHERAPY CARCINOMA RECTAL NEOPLASMS PROCTOSCOPY RADIOTHERAPY X-RAYS |