Author/Editor     Antolič, Vane; Pavlovčič, Vinko; Stanič, Uroš
Title     Anterior cruciate ligament function: evaluation with motion analysis system
Type     članek
Source     In: Magnificent milestones and emerging opportunities in medical engineering. Conference proceedings of the 19th annual international conference of the IEEE engineering in medicine and biology society; 1997 Oct 30-Nov 2; Chicago. Chicago: IEEE engineering,
Publication year     1997
Volume     str. 1742-5
Language     eng
Abstract     A new non-invasive quantitative method for evaluation of the anterior cruciate ligament function was tested in 24 asymptomatic subjects and in 37 patients with arthroscopically proven unilateral rupture of the anterior cruciate ligament using a 3-D optical motion analysis system. In the normal group, the average difference (d) in anterior displacement of the tibia between the left and the right side under a displacement force of 250N was 0.56 mm 0.56 mm (range 0.1 mm to 0.1 mm), while in patients the average difference (d) in anterior displacement of the tibia between the injured and uninjured side under a displacement force of 250N was 3.98 mm (range 0.6 mm to 7.9 mm). If the maximum value of the parameter d in normal subjects was used to discriminate between groups, 8% of the patients (3 patients) were classified as normal. On the other hand, if the minimum value of the parameter d in patients (0.6 mm) was used to discriminate between groups, 17% of the normal subjects (4 normal subjects) were classified as abnormal.
Descriptors     KNEE INJURIES
ANTERIOR CRUCIATE LIGAMENT
ARTHROSCOPY