Author/Editor     Poljak, Mario; Marin, Irena J; Seme, Katja; Brinovec, Vladimir; Matičič, Mojca; Meglič-Volkar, Jelka; Lešničar, Gorazd; Vince, Adriana
Title     Second-generation hybrid capture test and amplicor monitor test generate highly correlated hepatitis B virus DNA levels
Type     članek
Source     J Virol Methods
Vol. and No.     Letnik 97
Publication year     2001
Volume     str. 165-9
Language     eng
Abstract     The performance of the Digene Hybrid Capture II HBV DNA Test HC II and the Roche Cobas Amplicor Monitor Test (Cobas-HBV) was evaluated on 252 serum samples. One hundred and seventy-three samples were HBV DNA positive and 75 HBV DNA negative by both assays. Four samples were HBV DNA positive by Cobas-HBV only. Linear regression analysis showed that the HBV DNA concentrations obtained from both assays were significantly related (n=173, r=0.976, p<0.0001). The results of the study show that Hybrid capture II and Cobas-HBV could be used equally in the management for pstients with chronic HBV infection.
Descriptors     HEPATITIS B VIRUS
POLYMERASE CHAIN REACTION
DNA, VIRAL
HEPATITIS B