Author/Editor     Hall, GL; Wildhaber, JH; Černelč, M; Frey, U
Title     Evaluation of the interrupter technique in healthy, unsedated infants
Type     članek
Source     Eur Respir J
Vol. and No.     Letnik 18, št. 6
Publication year     2001
Volume     str. 982-8
Language     eng
Abstract     The interrupter technique may be used to monitor respiratory resistance and does not require active patient cooperation, but has yet to be applied in unsedated, spontaneously breathing infants. The aim of this study was to determine if the interrupter technique is feasible in spontaneously breathing infants and to investigate the influence of facemask types and analysis techniques on the interrupter resistance (Rint). Rint was measured in 14 healthy, unsedated, sleeping infants (aged 38.4 (31-56) days (mean (range)). Paired measurements were made using large volume, compliant (Mcomp) and small volume, rigid (Mrigid) facemasks. Flow and pressure were measured at the airway opening prior to- and following a brief airway occlusion (500 ms). Rint was calculated using four previously reported analysis techniques. Rint could be measured in all infants. Mcomp, independent of the analysis method significantly underestimated Rint (p<0.001). The variability and magnitude of Rint were significantly influenced by the choice of analysis method. The conclusion is that the interrupter technique is feasible in spontaneously breathing, unsedated infants. Equipment design and analysis method significantly influences interrupter resistance. Studies standardizing equipment and identifying the most appropriate analysis technique in this age group are needed.
Descriptors     AIRWAY RESISTANCE
TIDAL VOLUME
PEAK EXPIRATORY FLOW RATE
INFANT
SUPINE POSITION